Cookie Img
Send SMS Send Inquiry
Home » Products » Computed radiography » Kiran Cawo Digitize

Kiran Cawo Digitize

Kiran Cawo Digitize
Kiran Cawo Digitize
Product Code : COMR-01
Trade Information
  • Payment Terms
  • Cash Advance (CA), Cash in Advance (CID), Cheque
  • Main Export Market(s)
  • Asia, Australia, Central America, North America, South America, Eastern Europe, Western Europe, Middle East, Africa
  • Main Domestic Market
  • All India
Product Description

overview

The affordable, compact Kiran Digitizer 3.2 is designed as an ideal solution cost effective for digital Radiography without compromising on image quality.

Affordable computed radiography solution that makes no compromises on image quality.

Technical Specifications

Cassette SizesSpatial Resolution
35 x 43 cm (14 x 17”)

10 pixels/mm

(including for Full Leg / Full Spine)

24 x 30 cm (with cassette adapter)10 pixels / mm

GENERAL

Digitizer type

  • Single cassette feed
  • Throughput: UP to 78 plates/hour(depending on size)

Display

  • LED Status Indicator
  • Status and error messages on external PC monitor

Greyscale resolution

  • Data acquisition: 20 bits / pixel
  • Output to processor: 16 bits/pixel square root compressed

Dimensions and weight

  • W x D x H: 580 x 700 x 471 mm
  • Depth without cassette unit and extension: 380 mm
  • Weight: 29 kg

Power

  • Autoranging external poser supply (24 V output)
  • Input:

                •100 V - 240 V

                •< 2 A

                •50 / 60 Hz

                •Fuse: Europe max 16 A; USA max 15 A

Storage

  • Packed device shall withstand the following mechanical conditions: IEC 721-3-1: class 1M2 and IEC 721-3-2(1993): class 2M3; including sea transport
  • In line with IEC721-3-1: class 1K4

Minimum requirements

  • CR MD 1.0 General Plate
  • CR MD 1.0 General Cassette
  • CR MD 1.0 Cassette Adapter 24 x 30
  • CR MD 1.0 F General Plate
  • CR MD 1.0 F General Cassette
  • CR DD 1.0 Vet Plates
  • CR DD 1.0 Vet Cassette
  • NX

Environmental conditions

  • In line with: IEC 721-3-3 (1997): class 3K2,
  • Temperature: 15 - 35° C

Environmental effects

  • Noise level: max. 65 dB (A)
  • Heat dissipation: standby 30 W, max. 140 W

Mobile use

  • In line with IEC721-3-3 (1997): 3K2 with the following restrictions:
  • Temperature: +15° C to +35° C
  • Humidity: 15 - 75 % RH (non-condensing)
  • During transport with mobile kit: in line with
  • IEC721-3-5: 5K1 and 5M3

Transport

  • In line with: IEC 721-3-2 (1997): class 2K2, with the following restrictions: -25 to +55° C (-4 to 131° F)

Safety

  • Approvals
  • CE, cNRTLus

 


TRIVITRON HEALTHCARE PVT. LTD.

Contact Us

( *represents compulsory fields )

Tips on getting accurate quotes. Please include product name, order quantity, usage, special requests if any in your inquiry.



TRIVITRON HEALTHCARE PVT. LTD. All Rights Reserved.